| Transformers medical apps and insulation monitor
EN60601 and 93/42/EWG
| |||
![]() | ![]() | ![]() | ![]() |
| Download Product broschure |
| Electrical devices for applications connected to or placed close to patients must fulfill certain characteristics concerning maximum leakage current (<100µA in typical operation mode). All ALN/ ALE devices are also certified for use with conventional low-voltage system, outside health care areas. The devices comply to the following standards (approved according to the standards of LGA InterCert GmbH, Nuremberg:
All ALN/ALE devices already correspond to future requirements, since protective grounding is wired all the way from input to output plugs. For this reason ALN/ALE series, despite being characterized as insulating transformer, is approved for use medical appliances in health care environment.
A specially designed insulation monitor controls the insulation resistance (secundary winding). In case of lower limit failure the insulation monitor sets alarm acoustically and optically. Additionally the module serves as a transformer temperature monitor - possible overload or a lack of ventilation sets off the alarm as described above. |