1 Phase Isolating Tranformers

1AC Isolating transformer for use in IT networks / medical appliances according to VDE 0100-710MTS 3.15 - 8.0kVA according to IEC/EN61558-2-15

Isolating transformer for use in IT networks according to VDE 0100-710; device for medical appliances and health care environment. Primary-/ Secundary voltages: 230 / 230 V.
1 phase isolating transformer according to VDE 0100-710 requirements.
Construction according to protection class II; enhanced insulation (test voltage 4 KV between PRI / SEC shielding)
Static shielded winding between Primary- and Secundary winding; directed to insulated feeder clamp.

read more

1AC Isolating Transformer with Static Shielded Winding / Slim, space saving versionTTH 0.063 - 0.26kVA according to IEC/DIN EN61558-2-4

Extreme slim design for space-saving applicationsSpecial winding construction (1 chamber with shielded winding between PRI and SEC) allows excellent attenuation of mains failures (symmetric/ asymmetric) in SEC circuit.
Shielded winding delivery status: Common grounding connected to transformer core.
Standard voltages: PRI 230 V and SEC 230 V.
Voltage range PRI 100 V – 600 V and SEC 12 – 400 V on request.

read more

1AC Isolating Transformer with Static Shielded Winding / Slim design suitable for DIN rail mountingsTTHN 0.063 - 0.26kVA according to IEC/DIN EN61558-2-4

Extreme slim design for space-saving applicationsSpecial winding construction (1 chamber with shielded winding between PRI and SEC) allows excellent attenuation of mains failures (symmetric/ asymmetric) in SEC circuit.
Shielded winding delivery status: Common grounding connected to transformer core.
Standard voltages: PRI 230 V and SEC 230 V.
Voltage range PRI 100 V – 600 V and SEC 12 – 400 V on request.

read more